Everything else / Bulk

Cleaning Silica Gel Bonded HPLC Columns

The key to regenerating contaminated HPLC columns is to know the nature of the contaminants and to find the right solvent to remove. If contamination is caused by the accumulation of strongly retained materials during repeated injections, simple steps to remove these contaminants can often restore their chromatographic behavior.

Occasionally, after many operations, the column is rinsed with 90 to 100% solvent B (the stronger solvent in the dual solvent reversed phase system) to remove contaminants.

For example, the residual lipid in the column can be a nonaqueous solvent such as methanol, acetonitrile or tetrahydrofuran. If you are using a buffer system, do not switch directly to a strong solvent.

Sudden switching to a high concentration of organic solvent may cause buffering in the HPLC flow system, which can lead to larger problems such as plugging of the column, blockage of the pipe, pump Leakage, piston damage, or injection valve shaft failure.

The unbuffered mobile phase should be used first (i.e. the buffer should be replaced with water). Replace the strong solvent after rinsing 5 to 10 volumes.

EChroma C8 HPLC Cloumns

SiO2 HPLC Columns

Location : Xi'an City , Shaanxi Province, P.R.China, 710077 Xi'an,
Contact : Bruce Lu, +86 29 89284429

Good deal: buying from seller

Visit our guide for a secure transaction! Under no circumstances Algomtl can be held responsible for the content of the ads published. Only the sellers and buyers responsability is involved in case of dispute.
We invite you to read our terms of use. You can also visit our FAQ section and see our information section on the risks associated with counterfeiting.
This page is about importers and exporters of
Cleaning Silica Gel Bonded HPLC Columns

Search in a category : Everything else / Bulk
Search in a category :
silica, hplc, columns, bonded, cleaning
Hong Kong 999077 Equipment goods
Unitronics Samba

Saturday, May 04, 2024

Unitronics Samba – Automate Control Engineering Ltd. (winston@automate-ace.com) Still using text display? it's SAMBA time! Improve your competitive advantage. Great look - incredible price. System integrators and OEMs often develop applications that require PLC control, but that...

Automate Control Engineering Ltd.

Germany 31675 Furnitures
Wood pellets for sale

Saturday, November 21, 2015

Specification and quotation for our wood pellet best quality. PRODUCT NAME: WOOD PELLET Form: Stick Color: light yellow Type of wood: Pine, oak, beech and spruce. Diameter: 6-8 mm Length: 10-12 mm Ash: 0.6% max Humidity: 8% max Calories: 4200-4600KCAL / KG Inner density...

A & B IMEX

  • 31675 - Bückeburg
  • +49 (0)5722 2891328
Korea (Republic) 411-808 Health / Hygiene
Hematology Auto Slide Stainer

Saturday, September 15, 2012

Quantity : Minimum Or - Price : T/T in advance, L/C

This Hematology Auto Stainer provides fast, easy and stable staining process for staining blood according to Wright, Wright-Giemsa and May-Grünward staining method by automated system control. This new innovated stainer features advanced functions and performances over manual staining...

Dagatronics Corporation

  • 411-808 - Geonggi-do
  • +82 3 19 16 80 05