Normal Problem and Solution in GC Test
More than 90% of gas chromatographic problems occur in the inlet, and it is necessary to see which problems may occur in the inlet.
Problem No.1: Leak
If there is a leak, the pressure on the inlet may not be maintained, and the instrument will alarm. If the problem cannot be solved in time, the instrument will automatically turn off the pressure. Leakage is two-way. The carrier gas will leak out and the air will leak in. The air entering the chromatographic system, especially oxygen, is highly destructive, especially at high temperatures, where oxygen in the system can damage the stationary phase coating of the column. It results in much shorter of column life.
The main leak points as following:
1. Septum; the holes in the septum will become larger and larger after repeated use, and the seal will be problematic. For gas chromatography running around the clock, it is better to replace one septum per day.
2. O-ring outside the liner; long-term heating and cooling will also lead to rubber aging, sealing is reduced, O-rings need to be replaced in time.
3. Column seal; After repeated tightening, the sealing performance will also be reduced, try not to reuse it as many times as possible.
Problem No.2: Block
In general, the carrier gas and gas samples will not block the pipeline. Only when the sample contains some high-boiling compounds, some samples will condense in the split flow path due to untimely discharge, resulting in blockage of the pipeline or Partial blockage, which directly affects the accuracy and stability of the split ratio. The chromatographic results is that the repeatability of peak area is not good.
Problem No.3: Sample Residue
Although we have purged through the septum and shunted to minimize sample residue, there are many stubborn samples where the sample remains, including septa, liners, shunt plates, and even the front of the column. Replacing these consumables on the inlet can effectively solves the problem of sample residue. Selecting more inert liner and column also avoids sample residue better.
Contact : Bruce Lu, +86 29 89284429
Good deal: buying from seller
We invite you to read our terms of use. You can also visit our FAQ section and see our information section on the risks associated with counterfeiting.
|
This page is about importers and exporters of Normal Problem and Solution in GC Test Search in a category : Everything else / Bulk Search in a category : test, solution, normal, problem |
Friday, January 16, 2026
Action return pallets for wholesalers, professional clearance and European expor
Quantity : 30000 - Price : 0,50 €
We offer return pallets exclusively composed of products sourced from the Action retail chain, intended for wholesalers, clearance professionals, resellers, and exporters. These liquidation pallets contain non-food fast-moving consumer goods, carefully selected for their high resale potential and quick turnover. Offer details: - Return pallets from large retail distribution – Action brand - Approx. 1,000 items per pallet (80% of pallets) - Average price: around €0.50 per item...
TRANSIDI
- 22440 - PLOUFRAGAN
- 09 72 63 03 26
- 33(0)603670505
Tuesday, January 06, 2026
WT- 480T / R WT-480T with hidden electric condenser microphone or you can use lapel microphone. WT-480R with hidden super thin speaker, could be applied without earphone conditions. Match with dual slots charging stand HDC-202 or 30 slots portable charging HDC-30, easy to recharge battery...
Meicheng Audio Video Co., Ltd.
- meicheng
- 23511 - New Taipei City
- 886 2 82280311
Tuesday, January 06, 2026
Quantity : 99999 - Price : 1,00 €
Argile Rouge ARGANisme L'argile rouge est riche en fer et en oligo-éléments. De quoi favoriser la circulation sanguine et redonner de la luminosité à la peau. Elle nettoie et purifie également la peau en la débarrassant de ses impuretés. Idéale pour les peaux ternes ou sèches En...
ets-lachgar
- etslachgar
- 400000 - Marrakech
- +21 2 24 43 68 94
- +21 2 61 64 43 09