Normal Problem and Solution in GC Test
More than 90% of gas chromatographic problems occur in the inlet, and it is necessary to see which problems may occur in the inlet.
Problem No.1: Leak
If there is a leak, the pressure on the inlet may not be maintained, and the instrument will alarm. If the problem cannot be solved in time, the instrument will automatically turn off the pressure. Leakage is two-way. The carrier gas will leak out and the air will leak in. The air entering the chromatographic system, especially oxygen, is highly destructive, especially at high temperatures, where oxygen in the system can damage the stationary phase coating of the column. It results in much shorter of column life.
The main leak points as following:
1. Septum; the holes in the septum will become larger and larger after repeated use, and the seal will be problematic. For gas chromatography running around the clock, it is better to replace one septum per day.
2. O-ring outside the liner; long-term heating and cooling will also lead to rubber aging, sealing is reduced, O-rings need to be replaced in time.
3. Column seal; After repeated tightening, the sealing performance will also be reduced, try not to reuse it as many times as possible.
Problem No.2: Block
In general, the carrier gas and gas samples will not block the pipeline. Only when the sample contains some high-boiling compounds, some samples will condense in the split flow path due to untimely discharge, resulting in blockage of the pipeline or Partial blockage, which directly affects the accuracy and stability of the split ratio. The chromatographic results is that the repeatability of peak area is not good.
Problem No.3: Sample Residue
Although we have purged through the septum and shunted to minimize sample residue, there are many stubborn samples where the sample remains, including septa, liners, shunt plates, and even the front of the column. Replacing these consumables on the inlet can effectively solves the problem of sample residue. Selecting more inert liner and column also avoids sample residue better.
Contact : Bruce Lu, +86 29 89284429
Good deal: buying from seller
We invite you to read our terms of use. You can also visit our FAQ section and see our information section on the risks associated with counterfeiting.
|
This page is about importers and exporters of Normal Problem and Solution in GC Test Search in a category : Everything else / Bulk Search in a category : test, solution, normal, problem |
Friday, February 06, 2026
Action return pallets for wholesalers, professional clearance and European expor
Quantity : 30000 - Price : 0,50 €
We offer return pallets exclusively composed of products sourced from the Action retail chain, intended for wholesalers, clearance professionals, resellers, and exporters. These liquidation pallets contain non-food fast-moving consumer goods, carefully selected for their high resale potential and quick turnover. Offer details: - Return pallets from large retail distribution – Action brand - Approx. 1,000 items per pallet (80% of pallets) - Average price: around €0.50 per item...
TRANSIDI
- 22440 - PLOUFRAGAN
- 09 72 63 03 26
- 33(0)603670505
Wednesday, May 20, 2015
Quantity : 1000 - Price : N/A
HD4600 4 Ports HDMI Switch / 6 Ports Splitter Extender over CAT5e/6 It's the innovative and powerful solution for HDMI Extension. 4 HDMI sources are selectable and each of 6 RJ-45 ports cascade-chainable receiver which is capable of connecting another 2 receivers. The extended...
ACAFA Information Co.
- epanio
- 335 - Dasi Township
- +88 6 33 07 13 00
Monday, May 23, 2016
Quantity : 100000 met - Price : 0.280usd
Thomsun Industries LLC is a professional PPR manufacturers and ppr suppliers in Dubai UAE. T-Mech pressure piping system is considered to be the best carrier of potable water, using the revolutionary raw material of PPR-80 of Borouge brand. T-Mech PPR major products range covers...
Thomsun industries LLC
- Lily Kutty
- 6419 - Dubai
- +97 1 48 85 40 00
- +971551188058