Normal Problem and Solution in GC Test
More than 90% of gas chromatographic problems occur in the inlet, and it is necessary to see which problems may occur in the inlet.
Problem No.1: Leak
If there is a leak, the pressure on the inlet may not be maintained, and the instrument will alarm. If the problem cannot be solved in time, the instrument will automatically turn off the pressure. Leakage is two-way. The carrier gas will leak out and the air will leak in. The air entering the chromatographic system, especially oxygen, is highly destructive, especially at high temperatures, where oxygen in the system can damage the stationary phase coating of the column. It results in much shorter of column life.
The main leak points as following:
1. Septum; the holes in the septum will become larger and larger after repeated use, and the seal will be problematic. For gas chromatography running around the clock, it is better to replace one septum per day.
2. O-ring outside the liner; long-term heating and cooling will also lead to rubber aging, sealing is reduced, O-rings need to be replaced in time.
3. Column seal; After repeated tightening, the sealing performance will also be reduced, try not to reuse it as many times as possible.
Problem No.2: Block
In general, the carrier gas and gas samples will not block the pipeline. Only when the sample contains some high-boiling compounds, some samples will condense in the split flow path due to untimely discharge, resulting in blockage of the pipeline or Partial blockage, which directly affects the accuracy and stability of the split ratio. The chromatographic results is that the repeatability of peak area is not good.
Problem No.3: Sample Residue
Although we have purged through the septum and shunted to minimize sample residue, there are many stubborn samples where the sample remains, including septa, liners, shunt plates, and even the front of the column. Replacing these consumables on the inlet can effectively solves the problem of sample residue. Selecting more inert liner and column also avoids sample residue better.
Contact : Bruce Lu, +86 29 89284429
Good deal: buying from seller
We invite you to read our terms of use. You can also visit our FAQ section and see our information section on the risks associated with counterfeiting.
|
This page is about importers and exporters of Normal Problem and Solution in GC Test Search in a category : Everything else / Bulk Search in a category : solution, normal, test, problem |
Friday, June 03, 2016
Rassiya Sarl, is a trading company located in Casablanca, Morocco. It combines four different sectors of service, trading, transportation, and consulting , offering a vertical solution to the supplier and the wholesaler. The main activity of the company - wholesale supply of fresh vegetables...
Rassiya SARL
- 20000 - Casablanca
- +212 661 49 84 71
Wednesday, April 16, 2014
Quantity : 2880 - Price : 0,99 €
ORIGINAL ENERGY DRINK TENEUR ÉLEVÉE EN CAFÉINE(32 mg/100 ml) Boisson gazeuse avec Taurine et Caféine INGREDIENTS: Eau carbonique, Sucre, Glucose, Régulateur acide: Acide Citrique, Correcteur d'acidité: Citratede Sodium, Taurine (0,4%), Glucuronolactone (0,24%), Caféine (0,032%)...
FASHIONRICO FR BEVERAGE
- frdrinks
- 09160 - CAUMONT
- 09 72 30 91 95
- 06 95 44 22 14
Thursday, January 17, 2013
Quantity : 18MT - Price : 70,00 €
Below is the Quotation and our specification . Wood Pellet... ... ... ... ... ... ... ... ..95Euros Hardwood Charcoal,... ... ... ... ... ... . 100Euros Sunflower Husk Pellet, ... ... ... ... ..80Euros Firewood, ... ... ... ... ... ... ... ... ... 75Euros Briquettes... ... ...
NCF COMPANY
- 06070 - IZMIR
- 00905428841676