Igg/Igm Antibody Rapid Test Kit with CE Certification and Euro Medical White List |
Product name | Catalog # | |
KN002-24 | KN002-48 | |
OFR1ab / N Dual Nucleic Acid Detection Kit (RT-qPCR) | 24 Tests /Kit | 48 Tests /Kit |
Description:
Novel pneumonia (NCP) is a respiratory infection caused by a new type of virus that was first identified in late 2019.
This kit utilizes fluorescence quantitative probe-based PCR and guarantee a high specificity to ensure accurate one-step identification of ORF1ab and N genes. The kit offers a highly sensitive test with a limit of detection as low as 500 copies/ml. The results are available within 1.5 hour.
Kit Contents:
Product name | (48 Tests /Kit) |
Dual Enzyme solution A | 50 µl |
Dual Ncvo-O / N reaction solution A | 1 ml |
Dual Ncvo-O / N positive control A | 200 µl |
Dual Negative control A | 200 µl |
Note: The kit components of different lots can not be interchanged.
Storage and Transportation:
Store at -20ºC±5ºC. Can be stored at 4 ° C for a short period of time. Shelf life 1 year.
Repeated freezing and thawing ≤5 times.
Protect from light.
Applicable equipment:
ABI series, Bio-Rad, Roche series and other fluorescent real-time PCR instruments.
Specimens requirements:
Specimen type: nasopharyngeal swab from suspected infection; virus cell culture fluid, etc.
2. Specimen collection:
Pharyngeal, respiratory samples. Use a cotton swab to take secretions sample. Place the swabs in a centrifuge tube containing 3.0ml Virus preservation fluid and send them for inspection immediately.
Basic Protocol:
Take the test sample and extract the RNA nucleic acid according to the instructions of the nucleic acid extraction kit. The nucleic acid extraction product should be stored at -20 ° C.
According to the total number of test samples, the number of PCR reaction tubes needed is "N". N = number of samples + 1 negative control + 1 positive control.
The following protocol is recommended for a 20 µl reaction volume. If the volume of reaction changes, please adjust proportionally.
Components | Volume |
Dual Enzyme solution A | 1 µl |
Dual Ncvo-O / N reaction solution A | 19µl |
Final Volume: | 20 µl |
Add 5 µl each of the extracted RNA, positive and of negative control to corresponding reaction tubes. After assembling all the components, cover the tube caps and gently mix the contents of the tube, mix well, and centrifuge briefly.
Place the reaction tube inside a real-time PCR instrument.
Set the channel and sample information, reaction system volume 25 µl.
Select the following channels: FAM channel for ORF1ab, VIC channel for N.
Perform quantitative PCR using recommended cycling parameters settings:
Step | Temperature | Time | Number of cycles | Fluorescence Detector |
1 | 50°C | 15 min | 1 | Off |
2 | 95°C | 3 min | 1 | Off |
3 | 95°C | 10sec | 45 | Off |
60°C | 30 sec | On |
Result analysis:
1) Set the baseline: Generally, it is set to 6-15cycle for ABI 7500, 7700 and other instruments, 3-15cycle for PE5700, and 6-12cycle for MJ Research Option2. Under special circumstances, the baseline can be adjusted accordantly.
2) Set the threshold: The threshold line just exceeds the highest point of the negative control amplification curve (random noise line).
Result determination:
1) Positive: The Ct value of the two channels of the test sample is ≤ 38.0, and the curve has a significant exponential growth period, the result is determined to be positive;
2) Suspicious: Any test sample with a Ct value > 38.0 and < 40.0, repeat the experiment, if the Ct value is still < 40.0, and the curve has a significant exponential growth period, it is positive, otherwise it is negative;
3) Negative: If the Ct value is not detected or the Ct value is >40, the result is negative.
Quality control standards:
1) Negative quality control: no logarithmic growth phase or Ct value display in the amplification curve;
2) Positive quality control: the amplification curve has a significant logarithmic growth phase, and the Ct value is ≤30;
The above conditions should be met at the same time, otherwise the experiment is deemed invalid.
Limitations of detection method:
Precautions:
About AIVD Biotech
Established in 2014, AIVD Biotech Inc. is a biotech company, located in Shenzhen, China, a ISO 13485, ISO 9001 certificated enterprise. AIVD is a leading manufacturer of solutions for immunodiagnostics. Our company develops and produces protein raw materials: such as antibodies, antigens for immunoassay development; uncut sheets, POCT rapid tests kits, molecular diagnostics and offers various custom services for In Vitro diagnostic industry.
Good deal: buying from seller
This page is about importers and exporters of Igg/Igm Antibody Rapid Test Kit with CE Certification and Euro Medical White List Search in a category : Medical Search in a category : rapid, antibody, certification, test, white, euro, medical |
Friday, June 21, 2013
Quantity : 2000T/mont - Price : 880usd
99.5% high purity white fused alumina/WFA 3-5mm Description: White fused alumina oxide powder is a high-quality raw material, made of fused refin ing crystal, high purity, self-sharpening, good acid and alkali resistance, high temperature, thermal properties and stability it...
Saturday, July 18, 2015
Especificaciones: Tamaño de hoja: 210mm x 297mm, Internacional A4 Calidad: 100% importado de pulpa de madera virgen Blancura = 102-104%, blanco natural Capacidad: Copying100ppm alta velocidad, Laser Capaz, Especificaciones de chorro: Tamaño de hoja: 210mm x 297mm, Internacional A4 Calidad...
Monday, June 18, 2007
SARL REYNAERT IMPORT/EXPORT BUILDING MATERIALS/USTENSILS/METAL SCRAPS/YARD MACHINES We are selling many used CATERPILLAR yard machines from the UNITED STATES, CANADA or FRANCE , FOB or CIF prices. Any request must include a LOI with the name of your company, no curious please, no...